Features:
Meter
- Measurements displayed on a 3V2-digit LED display
- Overrange indicated by a flashing 1999
- The appropriate range selected automatically with the units of measure shown on the mode switch
- During resistance measurements, a separate LED illuminates when the current source has been activated
Equipment Current
- One range: 0.1 to 15.0 amps
- Accuracy: 5% of range
Voltage
- Three voltage ranges: 0.0 to 199.9 mV; 0 to 1999 mV; 0.0 to 199.9 V
AAMI Load used for these measurements
- Millivolt (mV) ranges autoranging
- Voltage (V) ranges used for line voltage measurements of the test receptacle powering the analyzer
- Millivolt (mV) ranges usable for external meter jack measurements
Test Receptacle
- Supplies power to the equipment under test
- 120 VAC at 15 amps maximum
- Push-button switches on front panel select
Power Requirements
- 117 VAC at 15 amps 50 to 60 Hz
- Uses very little power by itself (<100 mA)
- 15-amp rating for equipment under test plugged into the test receptacle
The DNI 232D Safety Analyzer is a precision voltage, current, and resistance meter designed expressly for testing the electrical safety level of both the patient environment and associated electrically operated equipment.
The 232D electrical safety analyzer measures grounding resistance of the equipment power cord; leakage current from the equipment's chassis or ground wire; equipment load current; voltage gradient (mV) and intergrounding resistance (m Ohms) using the external meter function; and power system AC voltages: neutral to hot, neutral to ground, and hot to ground.
The image is for reference only. See product specification for details.
Manufacturer | DNI Nevada |
---|---|
Condition | Used |