Additional Features:
- 5KV AC, 6KV DC Hipot Tester
- RS-232 interface standard for entry-level automation
- Patented SmartGFI® safety circuit protects the operator from shock hazards
- Patented VERI-CHEK® feature prompts users through steps to validate the instrument’s operation
- Patented CAL-ALERT® feature alerts the operator that the tester is due for re-calibration
- Built-in adjustable Continuity test for checking basic continuity
- Graphic LCD provides intuitive menu system to simplify the entire testing process from set-up to results
- Remote Safety Interlock feature prevents the high-voltage from being activated without the interlock enabled
- 10 Memories with 3 Steps per memory for storing and recalling test parameters
- PLC Remote Control for simple remote operation
- Interconnects with a HYAMP III Associated Research Ground Bond tester to form a complete test system
- Digitally controlled arc detection circuit allows the operator to program sensitivity levels for detecting arcs
- Minimum and maximum trip settings for safer and more accurate testing
- Comes complete with an adapter box for products terminated in a line cord
Input Specifications
- Voltage: 115/230 VAC ± 10%, user selectable
- Frequency: 50/60 Hz ± 5%
- Fuse: 3.15 A, fast acting 250 VAC
Dielectric Withstand Test Mode
- Output Rating 5000 V @ 20 mA AC; 6000 V @ 7.5 mA DC
- Voltage Setting Range: 0 - 5.00 KV AC; 0 - 6.00 KV DC
- Resolution: 0.01 KV
- Accuracy: ± (2% of setting + 5 V)
- Maximum Limit AC Range: 0.00 - 20.00 mA
- Resolution: 0.01 mA
- DC Range: 0 - 7500 μA
- Resolution: 1 μA
- Accuracy: AC and DC ± (2% of setting + 2 counts)
- Minimum Limit AC Range: 0.000 - 9.999 mA
- Resolution: 0.001 mA
- DC Range: 0.0 - 999.9 μA
- Resolution: 0.1 μA
- Accuracy: AC and DC ± (2% of setting + 2 counts)
- Arc Detection: Range: 0 - 9, 0 disabled
- Ground Fault Interrupt: GFI Trip Current: 450 μA max (AC or DC); HV Shut Down Speed: < 1ms
Current Display Auto Range
- AC Range 1: 0.000 - 3.500 mA
- Resolution: 0.001 mA
- Range 2: 3.00 - 20.00 mA
- Resolution: 0.01 mA
- DC Range 1: 0.0 μA - 350.0 μA
- Resolution: 1: 0.1 μA
- Range 2: 0.300 mA - 3.500 mA
- Resolution: 0.001 mA
- Range 3: 3.00 mA - 7.50 mA
- Resolution: 0.01 mA
- Accuracy: All Ranges ± (2% of reading + 2 counts)
- DC Output Ripple: ≤ 5% Ripple RMS at 6 KV DC @ 7.5 mA, Resistive Load
- Discharge Time: ≤ 200 ms
- The maximum capacitive load vs output voltage:
- 0.20 μF < 1 KV 0.050 μF < 4 KV
- 0.10 μF < 2 KV 0.040 μF < 5 KV
- 0.06 μF < 3 KV 0.015 μF < 6 KV
- AC Voltage Waveform: Sine Wave, Crest Factor = 1.3 - 1.5
- Output Frequency: Range: 50 or 60 HZ, User Selectable
- Output Voltage Regulation: ± (1% of output + 5 V) from no load to full load and over input voltage range.
Dwell Timer
- Range: AC 0, 0.3 - 999.9 sec (0 = Continuous); DC 0, 0.4 - 999.9 sec (0 = Continuous)
- Accuracy: ± (0.1% of reading + 0.05 sec)
Ramp Timer
- Range: Ramp-Up: 0.1 - 999.9 sec
- Ramp-Down: AC 0.0 - 999.9 sec
- DC 1.0 - 999.9 sec (0=OFF)
- Accuracy: ± (0.1% of reading + 0.05 sec)
- Ground Continuity Current: DC 0.1 A ± 0.01 A, fixed
- Ground Continuity Range: 0.0 Ω - 1.50 Ω
- Maximum Limit Resolution: 0.01 Ω
- Minimum Limit Accuracy: ± (3% of setting + 0.02 Ω)
- Ground Continuity Range: 0.0 Ω - 0.50 Ω
- Auto Offset Resolution: 0.01 Ω
- Accuracy: ± (3% of setting + 0.02 Ω)
Insulation Resistance Test Mode
Voltage Setting
- Range: 30 - 1000 VDC
- Resolution: 1 V
- Accuracy: ± (2% of setting + 5 V)
Resistance Display
- Range: 1 - 9999 MΩ (4 Digit, Auto Ranging)
- Resolution: 500 VDC - 1000 VDC
- 0.001: 1.000 - 9.999 MΩ
- 0.01: 10.00 - 99.99 MΩ
- 0.1: 100.0 - 999.9 MΩ
- 1: 1000 - 9999 MΩ
- Accuracy: ± (2% of reading + 2 counts) at test voltage; 500 - 1000 V and 1 - 999.9 MΩ
- ± (5% of reading + 2 counts) at test voltage
- 500 - 1000 V and 1000 - 9999 MΩ
- ± (8% of reading + 2 counts) at test voltage
- 30 - 500 V and 1 - 1000 MΩ
Maximum Limit
- Range: 0, 1 - 9999 MΩ (0=OFF)
- Resolution: 1 MΩ
- Accuracy: Same as Resistance Display
Minimum Limit
- Range: 1 - 9999 MΩ
- Resolution: 1 MΩ
- Accuracy: Same as Resistance Display
Ramp Timer
- Range: Ramp-Up: 0.1 - 999.9 sec
- Ramp-Down: 1.0 - 999.9 sec (0=OFF)
- Resolution: 0.1 sec
- Accuracy: ± (0.1% of reading + 0.05 sec)
Delay Timer
- Range: 0, 0.5 - 999.9 sec (0 = Continuous)
- Resolution: 0.1 sec
- Accuracy: ± (0.1% of reading + 0.05 sec)
- GFI Trip Current: 450 μA max
- HV Shut Down Speed: < 1 ms
General Specifications
- Mechanical: Bench or rack mount with tilt up feet
- Dimensions: (W x H x D) 8.46 x 3.5 x 14.57 in.; (215 x 89 x 370 mm)
- Interface: RS-232 interface standard for entry-level automation
- Memory: 10 Memories, 3 steps per memory