Features:

  • Durable PRTs with temperature range to 788°F (420°C) and accuracy to 0.025°C
  • Vibration and shock resistant
  • NVLAP-accredited calibration included, lab code 200706-0

Accessories (Included):

  • Fluke 5627A-6-G Precision Thermometer/RTD Temperature Probe with Gold Pins, 6 x 0.19in, -200 to 300°C

Specifications:

  • Resistance: Nominal 100Ω
  • Temperature Coefficient: 0.00385Ω/Ω/°C nominal
  • Temperature Range: -328 to 788°F (-200 to 420°C) (5627A-6 and 5627A-9 to 300°C; transition and cable temperature: 0 to 150°C)
  • Drift Rate (k=2): ±0.04°C at 0°C after 100 hours at 420°C
  • Sheath Material: 316 stainless steel
  • Leads: PTFE-insulated, nickel-plated stranded copper, 22AWG
  • Time Constant: 4 seconds maximum for 63.2% response to step change in water moving at 3fps
  • Bending Radius: Sheath may be ordered with a bend on a minimum radius of 0.75in (19mm) except for 2in (50mm) area of sheath near tip; Fluke lab requires 8in (20cm) of unbent sheath to re-calibrate
  • Calibration: Includes manufacturer's NVLAP-accredited (lab code 200706-0) calibration and table with R vs T values in 1°C increments from -196 to 500°C (to 300°C for 5627A-6 and 5627A-9) ITS-90 coefficients included
  • Immersion: At least 4in (100mm) recommended
  • Calibrated Accuracy: (Includes Calibration Uncertainty and 100-Hour Drift) (k=2) ±0.026°C at -196°C, ±0.046°C at 0°C, ±0.077°C at 200°C, ±0.124°C at 420°C
  • Size: 6 x 0.19in (152 x 4.7mm)

The 5627A-6-G is a probe from Fluke. Featuring a coil suspension element, this precision thermometer/RTD temperature probe delivers a design that's shock and vibration resistant. You can use your DIN/IEC RTD meter fully, since the unit conforms to standard 385 curves. With a mineral-insulated sheath that measures 9in x 0.19in and INFO-CON connections for 914X models, this thermometer provides a highly flexible design.Featuring a coil suspension element, this precision thermometer/RTD temperature probe delivers a design that? shock and vibration resistant. You can use your DIN/IEC RTD meter fully, since the unit conforms to standard 385 curves. With a mineral-insulated sheath that measures 6in x 0.19in and gold pin connections, this thermometer provides a highly flexible design.