Additional Features:
- Frequency range: 0.01-50 GHz
- +10 dBm calibrated output power
- -50 dBc harmonics <26.5 GHz typical
- SSB phase noise <-80 dBc at 10 GHz and 10 kHz offset
- Complete analog sweeper
- 1 Hz frequency resolution (Option 008)
- Pulse, amplitude and frequency modulation (Agilent 8360B Series only)
- -110 dBm output power with step attenuator (Option 001)
The Agilent 83650B is the standard of excellence for applications requiring the high performance and accuracy of a synthesized source and the speed and versatility of a sweep oscillator. Also called a sweeper, the unit has synthesized broadband frequency coverage and precise 1 Hz frequency resolution (Opt 008) are generated by indirect synthesis, enabling it to achieve low single-sideband phase noise performance. The HP 83650B also delivers excellent harmonic performance with harmonics at least 50 dB below the carrier from 2.0 to 26.5 GHz and at least 40 dB below the carrier above 26.5 GHz. This unit is compatible with Agilent scalar and vector network analyzers, noise figure systems, and millimeter wave source modules. Standard Commands for Programmable Instruments (SCPI) ensures compatibility with other HP equipment.
More Information:
Frequency Bands (CW Signals):
10 MHz to <2 GHz
2 GHz to <7 GHz
7 GHz to <13.5 GHz
13.5 GHz to <20 GHz
20 GHz to <26.5 GHz
26.5 GHz to <38 GHz
38 GHz to 50 GHz
Internal 10 MHz time base:
Accuracy = Calibration ± Aging Rate ± Temperature Effects ± Line Voltage Effects
Aging Rate: 5 10–10/day, 1 10–7/year With Temperature: 1 10–10/°C, typical
With Line Voltage: 5 10–10 for line voltage change of 10%, typical
Sweep Functions:
Control: Start/stop, center/span, marker (M1-M2), alternate sweep
Trigger: auto, external, single, or GPIB
CW and Manual Sweep Modes:
Accuracy: Same as time base
Switching Time:
- For steps within a frequency band:
- 15 ms + (step size/1 GHz) × 5 ms step size Maximum, or across band switch points: 50 ms
- Step or list modes within a frequency band:
- 5 ms2 + (step size /1 GHz) × 5 ms
- Accuracy: Same as time base
- Minimum Step Size: Same as frequency resolution
- Number of Points: 2 to 801
- Switching Time: Same as CW
- Dwell Time: 100 μs to 3.2 s
- Accuracy: Same as time base
- Minimum step size: Same as frequency resolution
- Number of points: 1 to 801
- Switching time: Same as CW
- Dwell time: 100 μs to 3.2 s
- Accuracy (sweep time >=100 ms and <=5 s):
-
Sweep widths >n × 10 MHz: (start, stop, and bandedge frequencies are phase-locked
corrected during sweep) Lesser of 1% of sweep width or n × 1 MHz + 0.1% of sweep width. - Sweep time:
-
10 ms to 100 s, 300 MHz/ms maximum rate
- Output frequencies <= 26.5 GHz
- Standard: +10
- Option 006: +10
- Output Frequencies >= 26.5 GHz and <= 40 GHz
- Standard: +5
- Option 006: +5
- Output Frequencies >= 40 GHz
- Standard: +2.5
- Option 006: +2.5
With attenuator (Option 001): Minimum settable output power is –110 dBm. Maximum leveled
output power is reduced by 1.5 dB to 20 GHz, 2 dB above 20 GHz, and 2.5 dB above 40 GHz.
Manufacturer | Agilent, HP, Keysight |
---|---|
Condition | Used |
Frequency | 50 GHz |