9009-B-156 Fluke Temperature Calibrator

New
$8,483.00
Free Shipping
 

Features: 

  • Two independently controlled temperature blocks, one for heating and one for cooling with a combined range of -15 to 350°C
  • Unique dual unit allows you to calibrate your sensor in the heated well while the cool well stabilizes; this eliminates the need to wait for multiple temperature points during calibration
  • Hot block provides temperatures from 50 to 350°C (122 to 662°F), while the cold block covers the range of -15 to 110°C (5 to 230°F)
  • Each block is controlled by a precision Hart Scientific temperature controller
  • Cold block is calibrated to within ±0.2°C with stability of ±0.05°C
  • Hot block's display is accurate to ±0.6°C with stability of ±0.05°C
  • NIST-traceable calibration is included for each of the two test blocks
  • Each temperature block includes two wells with removable inserts; calibrate four probes at once, or calibrate two probes at the same time with an external reference, or use the two temperature wells to get quick "zero" and "span" references for transmitter calibrations
  • Inserts are available to accommodate sensors of any size from 1/16" (1.6 mm) to 7/16" (11.1 mm)
  • RS-232 connection and Interface-it software allows you to control and monitor temperatures from your PC
  • For completely automated calibrations, Hart's MET/TEMP II software also integrates with the 9009
  • A single AC connection powers both temperature wells for convenience at the job site
  • Housed in a tough Pelican™ case that is both airtight and watertight
  • Both wells accept the same removable inserts so that you only have to keep one set on hand for both wells (four inserts are stored in the wells, and an additional four can be stored in a convenient holding rack inside the case)

Accessories:

  • Fluke 9009-Y-156 Dual Range Dry-Block Calibrator, Yellow

Applications:

  • Ideal for the following applications: Laboratory, Education, R&D, Calibration Labs, Field Calibration

Specifications:

  • Temperature Range - Cold Block: -15 to 110°C (5 to 230°F) (-8°C [18°F] w/ hot block @ 350°C [662°F])
  • Hot Block: 50 to 350°C?(122 to 662°F)
  • Accuracy - Cold Block: ±0.2°C from 1/N to 200°C
  • Hot Block: ±0.6°C from 200 to 350°C
  • Stability - ±0.05°C
  • Well-to-Well Uniformity - ±0.1°C
  • Well Depth - 4" (102 mm)
  • Stabilization Time - 8 mins
  • Heating Time - Cold Block: 15 mins from 25 to 110°C
  • Hot Block: 30 mins from 25 to 350°C
  • Cooling Time - Cold Block: 16 mins from 25 to -15°C
  • Hot Block: 40 mins from 350 to 100°C
  • Removable Inserts - Two 6.4 mm (1/4") and two 4.8 mm (3/16") inserts

The Fluke 9009-B-156 is a temperature calibrator. This unit has two independently controlled temperature blocks, one for heating and one for cooling with a combined range of -15 to 350°C.

More Information
ManufacturerFluke
ConditionNew
Looking for other manuals? Search all of our available manuals here.