Additional Features:
DC Voltage
- Amplitude: ±1mV to ±200V into 1MΩ; ±1mV to ±5V into 50Ω
- Accuracy: ± (0.025% + 25μV)
- Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
- Deviation: ±11.2%
Squarewave
Amplitude
- Range: 40μV to 200V pk-pk into 1MΩ; 40μV to 5V pk-pk into 50Ω
- Polarity: Positive, negative or symmetrical about ground
- Accuracy (10Hz to 10kHz): <1mV ± (1% + 10μV); 1mV-21mV ±(0.10% + 20μV); 21mV-556mV ±(0.10% + 1μV); 556mV-210V ±(0.05% + 1μV)
- Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
- Deviation: ±11.2%
- Rise/Fall Time: <100V <150ns; 100V <200ns
- Aberrations: <2% peak for first 500ns
- Frequency Range: 10Hz to 100kHz
- Accuracy: ±10ppm (±0.25ppm with Option 100)
- Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
Low-Edge Pulse
Amplitude
- Range: 5mV to 3V pk-pk into 50Ω
- Accuracy: ±3%
- Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
- Deviation: ±11.2%
- Rise/Fall Time: 500ps return to ground
- Mark/Space Ratio: 1:9
- Aberrations: <2% peak for first 10ns; <0.25% peak 10 ns to 1μs; <0.1% peak beyond 1μs
- Frequency Range: 10Hz to 2MHz
- Accuracy: ±10ppm (±0.25ppm with Option 100)
- Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
High-Edge Pulse
Amplitude
- Range: 1V to 200V pk-pk into 1M½
- 1V to 5V pk-pk into 50½
- Accuracy: ±3%
- Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
- Deviation: ±11.2%
- Rise/Fall Time: <100V <150ns; >=100V <200ns
- Mark/Space Ratio: 1:1
- Aberrations: <2% peak for first 500ns; <0.1% peak 500ns to 100μs; <0.01% peak beyond 100μs
- Frequency: 10Hz to 100kHz
- Accuracy: ±10ppm (±0.25ppm with Option 100)
- Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
Fast-Edge (only available on Model 9520 and Model 9530 Active Heads)
Amplitude
- Range: 5mV to 3V pk-pk into 50Ω
- Accuracy: ±3%
- Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
- Deviation: ±11.2%
- Rise/Fall Time: 150ps return to ground
- Mark/Space Ratio: 1:9
- Aberrations: <3% peak for first 1ns; <2% peak 1 ns to 10ns; <0.25% peak 10 ns to 50ns
- Frequency:
- Range: 10Hz to 2MHz
- Accuracy: ±10ppm (±0.25ppm with Option 100)
- Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
Timing Markers
- Styles: Square/Sine, Pulse or Narrow Triangle
- Square/Sine Period Square: 10ns to 50s
Period Sine
- 0.5ns to 10ns
Pulse
- Period: 1μs to 50s
- Rise/Fall Time: <700ps
Narrow Triangle
- Period: 1μs to 50s
- Rise/Fall Time: 2.5% of period
- Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous for period of all waveshapes
Timing Accuracy:
- Normal: ±10ppm
- With Option 100: ±0.25ppm
- Timing Jitter: ≤10ps pk-pk
- Deviation: ±45% for period
- Amplitude: 100mV to 1V pk-pk
- Sub-Division: Every10th marker can be set to higher amplitude for periods ≥1μs for all waveshapes
Leveled Sine and Dual Sine
- Frequency Range: 0.1 Hz to 3.2 GHz
Accuracy:
- Normal: ±12ppm
- With Option 100: ±0.25ppm for f ≥12kHz; ±3ppm max for f < 12kHz
- Deviation: ±11.2%
Amplitude (Leveled Sine into 50Ω):
- 0.1Hz - 550MHz 4.44mV to 5.560V pk-pk
- 550MHz - 2.5GHz 4.44mV to 3.336V pk-pk
- 2.5GHz - 3.2GHz 4.44mV to 2.224V pk-pk
- Accuracy ±1.5% at 50 kHz
Flatness (Leveled Sine relative to 50kHz):
- 0.1Hz - 100MHz ±1.5%
- 100MHz - 550MHz ±3%
- 550MHz - 1.1GHz ±4%
- 1.1GHz - 3.2GHz ±5%
- Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
Sine Purity
- 2nd Harmonic <-35dBc
- 3rd Harmonic <-40dBc
- All Other Spurious
- Signals <-40dBc (typical)
Input Impedance
Resistance Measurement
- Range: 10Ω-150Ωand 50kΩ-12MΩ
Accuracy:
- Ω10-40 ±0.5%
- 40 - 90 ±0.1%
- 90-150 ±0.5%
- 50k - 800k ±0.5%
- 800k - 1.2M ±0.1%
- 1.2M -12M ±0.5%
Capacitance Measurement:
- Range: 1pF to 95pF
- Accuracy: 1pF - 35pF 2% ± 0.25pF; 35pF - 95pF 3% ± 0.25pF
Current
Amplitude:
- DC: ±100μA to ±100mA
- Squarewave: 100μA to 100mA pk-pk
- Accuracy: ± (0.25% + 0.5μA)
- Frequency: 10Hz to 100kHz
- Accuracy: ±10ppm (±0.25ppm with Option 100)
- Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
Composite Video Output
- Amplitude: 1.0V, 0.7V, 0.3V
- Pattern: White, Grey or Black
- Sync Polarity: Positive or negative
- Standards: 625-line 50Hz or 525-line 60Hz
LF Linear Ramp
- Waveforms: 1V pk-pk symmetrical triangle
- Ramp Time: 1ms to 1s
Overload Pulse
- Amplitude: 5V to 20V into 50½
- Polarity: Positive or negative
- Duration: 0.2s to 100s
- Trigger: Manual
Zero Skew
- Unadjusted Skew: ±50ps channel to channel
- Adjusted Skew: ±5ps channel to channel
- Frequency Range: 10Hz to 100MHz
Auxiliary Input
- Signal Routing: Rear input to any Active Head
Maximum Input:
- Voltage: ±40V pk-pk
- Current: ±400mA pk-pk
Trigger
- Amplitude: ≥1V pk-pk into 50Ω
- Risetime: <700ps
Rate:
- User Selectable: f (up to 120 MHz), f/10 or f/100
- Free Run: 100Hz
Reference Frequency Input
- Frequency Range: 1MHz to 20MHz in 1MHz steps
- Level: 90 mV to 1V pk-pk (typical)
- Lock Range: ±50ppm
Reference Frequency Output
- Frequency: 1MHz or 10MHz
Level:
- Into 50Ω: 1V pk-pk (typical)
- Into 1MΩ: 2V pk-pk (typical)
Environment
Temperature:
- Operating: 5°C to 40°C
- Storage: 0°C to 50°C
- Humidity: (non-condensing)
- Operating: <90% over 5°C to 30°C
- <75% over 30°C to 40°C
- Storage: <95% over 0°C to 50°C
Power
- Voltage: 95V to 132V rms or 209V to 264V rms
- Frequency: 48Hz to 63Hz
- Consumption: 400 VA
- Warm-up: 20 minutes
Safety
- Designed to UL3111 and EN61010-1-1:1993/A2:1995
- CE Marked
EMC (including options)
- Emissions: EN55011:1991
- Immunity: EN50082-1:1992
- FCC Rules part 15 sub-part J class B
The WaveTek Model 9500-3200 is a state-of-the-art calibrator offering oscilloscope test and calibration capabilities from a single source, providing wide functionality.
Manufacturer | WaveTek |
---|---|
Condition | Used |
100 | High-Stability Crystal Reference |
---|---|
5 | 5-Channel Output (allows any mix of 9510/9520/9530 Heads up to a total of five. Upper frequency limited by Model 9500 mainframe.) |
9530 X 5 | Five (5) 3.2 GHz Active Heads with 150 ps and 500 ps pulse risetime |