Additional Features:
- AC Output to 5000 volts; DC output to 7200 volts delivers up to 20 milliamps. Custom configurations up to 300 mA available
- Leakage current 0-10 mA AC; 0-5 mA DC
- AC/DC selectable. Logic controlled, one button test. LEDÕs indicate test progress. Upon failure, buzzer sounds and testing is terminated
- Leakage current limit, voltage ramp-up time and test duration are controlled from the rear panel. Output voltage adjustable from front and rear
- Automatic voltage rampdown circuit eliminates residual voltage
- Designed for product development and laboratory use
- High voltage test jack output
- Voltage ramp, shutdown on failure and test timer are defeatable
- Operation instructions provided on the rear panel
- Ergonomically designed for safety, speed and efficiency
- Calibration resistor mounted on rear panel
- One year calibration cycle
- Economical, durable and compact
The Compliance West HT-5000 is best suited for engineers conducting type testing and product design. Front panel voltage dial-up and AC/DC rocker switch make this unit easy to use. Test termination on failure, voltage ramp time, and the test timer may be disabled for specific research.