M150AC Rod L Electronics HiPot

Used

Additional Features:

  • Safe, Fast, Efficient
  • Fully programmable test parameters
  • Digital Display
  • Resolution of programmable test parameters is 0.5%
  • Performs an AC dielectric strength (Hipot) test concurrent with a ground continuity test
  • Applies a test potential of up to 5000 VAC at 50 milliamps (user programmable)
  • Performs tests in full compliance with UL, VOE, BSI, IEC, CSA, and other test standards

Specifications

  • Output Test Voltage: 100 VAC-5K VAC*
  • Output Voltage Ramp Time: 50 V/sec to 5kV/sec*
  • Total Current Trip Point: 50 mA (maximum)*
  • Real Current Trip Point: 5 mA (maximum)*
  • Under Current Trip Point: 5 mA (maximum)*
  • Test Time: 1 sec - 100 sec*
  • *All of the above are User Programmable
  • Shutdown Time: 2 milliseconds
  • Arc Detection: Arc duration 1OµS or greater
  • Ground Continuity: Low current (1.5 Amp @ 1.5V)
  • Input Power Required: 115/230 VAC ± 10%, 47-63 Hz, 250 watts maximum, 30 watts typical
  • Dimensions: 16.75" x 7.00" x 18.38" (43cm x 18cm  x 47cm)

IEEE-488 BUS-compatible - enables data logging during each test, and remote programming of:

  • Test Voltage
  • Test Time
  • Voltage Ramp Rate
  • Real Current Trip Point
  • Total Current Trip Point
  • Under Current Trip Point

Detects and indicates the following events:

  • Test Ready (Security Chassis Ground Sensed)
  • Test in Progress
  • Test Passed
  • Test Failed:
  • Arcing
  • Excessive Total Leakage Current
  • Excessive Resistive Leakage Current
  • Under Current
  • Lack of adequate Chassis Ground

Superior Safety Features

  • Visual and Audible alert indicating Hipot test in progress
  • Low Current Security Chassis Ground Circuit (ensures Device Under Test has ground connection of o.sn between chassis and power cord ground pin)
  • Fast HV shutdown within 2 milliseconds of HV test automatically on test failure or on command

IEEE-488 BUS Programming

The Model M150AC Hipot Test System, when operated via the Model M1088C IEEE-488 BUS INTERFACE, functions as both a " Talker "(T) and a "Listener" (L) in conjunction with the following signals:

  • Output Voltage (T/L)
  • Output Voltage Ramp Rate (L)
  • Test Time (L)
  • Total Current Trip Point (L)
  • Total Current Magnitude (T)
  • Real Current Trip Point (L)
  • Real Current Magnitude (T)
  • Under Current Trip Point (L)
  • Test Status: Ready (T); In -Progress (T); Passed (T); Failed (T)
  • Type of Failure (T)
  • Start (L)
  • Stop (L)
  • Reset (L)

These signals can also be controlled using any controller capable of inputting and outputting digital and analog signals.