Additional Features:
- Safe, Fast, Efficient
- Fully programmable test parameters
- Digital Display
- Resolution of programmable test parameters is 0.5%
- Performs an AC dielectric strength (Hipot) test concurrent with a ground continuity test
- Applies a test potential of up to 5000 VAC at 50 milliamps (user programmable)
- Performs tests in full compliance with UL, VOE, BSI, IEC, CSA, and other test standards
Specifications
- Output Test Voltage: 100 VAC-5K VAC*
- Output Voltage Ramp Time: 50 V/sec to 5kV/sec*
- Total Current Trip Point: 50 mA (maximum)*
- Real Current Trip Point: 5 mA (maximum)*
- Under Current Trip Point: 5 mA (maximum)*
- Test Time: 1 sec - 100 sec*
- *All of the above are User Programmable
- Shutdown Time: 2 milliseconds
- Arc Detection: Arc duration 1OµS or greater
- Ground Continuity: Low current (1.5 Amp @ 1.5V)
- Input Power Required: 115/230 VAC ± 10%, 47-63 Hz, 250 watts maximum, 30 watts typical
- Dimensions: 16.75" x 7.00" x 18.38" (43cm x 18cm x 47cm)
IEEE-488 BUS-compatible - enables data logging during each test, and remote programming of:
- Test Voltage
- Test Time
- Voltage Ramp Rate
- Real Current Trip Point
- Total Current Trip Point
- Under Current Trip Point
Detects and indicates the following events:
- Test Ready (Security Chassis Ground Sensed)
- Test in Progress
- Test Passed
- Test Failed:
- Arcing
- Excessive Total Leakage Current
- Excessive Resistive Leakage Current
- Under Current
- Lack of adequate Chassis Ground
Superior Safety Features
- Visual and Audible alert indicating Hipot test in progress
- Low Current Security Chassis Ground Circuit (ensures Device Under Test has ground connection of o.sn between chassis and power cord ground pin)
- Fast HV shutdown within 2 milliseconds of HV test automatically on test failure or on command
IEEE-488 BUS Programming
The Model M150AC Hipot Test System, when operated via the Model M1088C IEEE-488 BUS INTERFACE, functions as both a " Talker "(T) and a "Listener" (L) in conjunction with the following signals:
- Output Voltage (T/L)
- Output Voltage Ramp Rate (L)
- Test Time (L)
- Total Current Trip Point (L)
- Total Current Magnitude (T)
- Real Current Trip Point (L)
- Real Current Magnitude (T)
- Under Current Trip Point (L)
- Test Status: Ready (T); In -Progress (T); Passed (T); Failed (T)
- Type of Failure (T)
- Start (L)
- Stop (L)
- Reset (L)
These signals can also be controlled using any controller capable of inputting and outputting digital and analog signals.
Manufacturer | Rod L Electronics |
---|---|
Condition | Used |
No manuals are currently available. Search all of our available manuals here.