Features
- Balance budget & performance to address parametric testing of microwave components & receivers
- Perform LO up conversion for microwave backhaul links or CW blocking for receiver testing
- Maximize test throughput with 600-µs frequency switching
- Characterize microwave filters & amplifiers with the best combination of output power, low harmonics & full step attenuation
- Use as a high-stability system reference: the standard high-performance OCXO has an aging rate of 5x10-10 parts/day
Specifications
- Internal time base reference oscillator aging rate: 1 < ± 1 x 10^-7/year 2, < ± 5 x 10^-10/day after 30 days
- Initial achievable calibration accuracy: ± 4 x 10^-8 or ± 40 ppb
- Adjustment resolution: < 1 x 10^-10 (nom)
- Temperature effects: < ± 2 x 10^-8 from 20 to 30 °C (nom)
- Line voltage effects: < ± 1 x 10^-9 for ± 10% change (nom)
Reference output
- Frequency: 10 MHz
- Amplitude: ≥ +4 dBm, (nom) into 50 Ω load
External reference input
- Input frequency standard: 10 MHz
- Input frequency Option 1ER: 1 to 50 MHz (in multiples of 0.1 Hz)
- Lock range: ± 1 ppm (nom)
- Amplitude: 5 dBm ± 2 dB (nom) 3
- Impedance: 50 Ω (nom)
- Waveform: Sine or square
- Stability: Follows the stability of external reference input signal
The N5173B is a 40 GHz rf generator from Agilent Keysight HP.
Manufacturer | Agilent, HP, Keysight |
---|---|
Condition | Used |
Frequency | 40 GHz |
Looking for other manuals? Search all of our available manuals here.