3605 Associated Research HiPot

Used

Additional Features:

  • 5KV AC Hipot Tester
  • Patented SmartGFI® safety circuit protects the operator from shock hazards
  • Patented VERI-CHEK® feature prompts users through steps to validate the instrument’s operation
  • Patented CAL-ALERT® feature alerts the operator that the tester is due for re-calibration
  • Built-in adjustable Continuity test for checking basic continuity
  • Graphic LCD provides intuitive menu system to simplify the entire testing process from set-up to results
  • PLC Remote Control allows operators to remotely control the Hipot tester
  • Remote Safety Interlock feature prevents the high-voltage from being activated without the interlock enabled
  • Memories with 3 Steps per memory for storing and recalling test parameters
  • Electronic Ramp and Dwell timers for more consistent and reliable testing
  • Interconnects with an Associated Research Ground Bond tester to form a complete test system
  • Digitally controlled arc detection circuit allows the operator to program sensitivity levels for detecting arcs
  • Minimum and maximum trip settings for safer and more accurate testing
  • Comes complete with an adapter box for products terminated in a line cord

Input Specifications

  • Voltage: 115/230 VAC ± 10%, user selectable
  • Frequency: 50/60 Hz ± 5%
  • Fuse: 3.15 A, fast acting 250 VAC

Dielectric Withstand Test Mode

Output Rating 

  • 5000 V @ 20 mA AC; 6000 V @ 7.5 mA DC

Voltage Setting 

  • Range: 0 - 5.00 KV AC; 0 - 6.00 KV DC
  • Resolution: 0.01 KV
  • Accuracy: ± (2% of setting + 5 V)

Maximum Limit

  • AC Range: 0.00 - 20.00 mA
  • Resolution: 0.01 mA
  • DC Range: 0 - 7500 μA
  • Resolution: 1 μA
  • Accuracy: AC and DC ± (2% of setting + 2 counts)

Minimum Limit 

  • AC Range: 0.000 - 9.999 mA
  • Resolution: 0.001 mA
  • DC Range: 0.0 - 999.9 μA
  • Resolution: 0.1 μA
  • Accuracy: AC and DC ± (2% of setting + 2 counts)

Arc Detection

  • Range: 0 - 9, 0 disabled

Ground Fault Interrupt 

  • GFI Trip Current: 450 μA max (AC or DC)
  • HV Shut Down Speed: < 1ms

Current Display Auto Range

  • AC Range 1: 0.000 - 3.500 mA
  • Range 2: 3.00 - 20.00 mA
  • DC Range 1: 0.0 μA - 350.0 μA
  • Range 2: 0.300 mA - 3.500 mA
  • Range 2: 3.00 mA - 7.50 mA
  • Accuracy All Ranges ± (2% of reading + 2 counts)

DC Output Ripple 

  • ≤5% Ripple RMS at 6 KV DC @ 7.5 mA, Resistive Load

Discharge Time

  • <=200 ms

The maximum capacitive load vs output voltage:

  • 0.20 μF < 1 KV 0.050 μF < 4 KV
  • 0.10 μF < 2 KV 0.040 μF < 5 KV
  • 0.06 μF < 3 KV 0.015 μF < 6 KV

AC Voltage Waveform

  • Sine Wave, Crest Factor = 1.3 - 1.5, <2% THD

Output Frequency

  • Range: 50 or 60 HZ, User Selectable

Output Voltage Regulation

  • ± (1% of output + 5 V) from no load to full load and over input voltage range.

Dwell Timer 

  • Range: AC 0, 0.3 - 999.9 sec (0 = Constant)
  • DC 0, 0.4 - 999.9 sec (0 = Constant)
  • Accuracy: ± (0.1% of reading + 0.05 sec)

Ramp Timer 

  • Range: Ramp-Up: 0.1 - 999.9 sec
  • Ramp-Down: AC 0.0 - 999.9 sec
  • DC 1.0 - 999.9 sec (0=OFF)
  • Accuracy: ± (0.1% of reading + 0.05 sec)

Ground Continuity Current

  • DC 0.1 A ± 0.01 A, fixed

Ground Continuity Maximum Limit, Minimum Limit

  • Range: 0.0 Ω - 1.50 Ω
  • Resolution: 0.01 Ω
  • Accuracy: ± (3% of setting + 0.02 Ω)

Ground Continuity Auto Offset

  • Range: 0.0 Ω - 0.50 Ω
  • Resolution: 0.01 Ω
  • Accuracy: ± (3% of setting + 0.02 Ω)

Insulation Resistance Test Mode

Voltage Setting

  • Range: 30 - 1000 VDC
  • Accuracy: ± (2% of setting + 5 V)

Resistance Display

  • Range: 1 - 9999 MΩ (4 Digit, Auto Ranging)
  • Resolution: 500 VDC - 1000 VDC

MΩ                MΩ

0.001      1.000 - 9.999

0.01        10.00 - 99.99

0.1          100.0 - 999.9

1              1000 - 9999

Accuracy

  • ± (2% of reading +2 counts) at test voltage; 500 - 1000 V and 1 - 999.9 MΩ
  • ± (5% of reading +2 counts) at test voltage; 500 - 1000 V and 1000 - 9999 MΩ
  • ± (8% of reading +2 counts) at test voltage; 30 - 500 V and 1 - 1000 MΩ

Maximum Limit 

  • Range: 0, 1 - 9999 MΩ (0=OFF)
  • Accuracy: Same as Resistance Display

Minimum Limit 

  • Range: 1 - 9999 MΩ
  • Accuracy: Same as Resistance Display

Ramp Timer 

  • Range: Ramp-Up: 0.1 - 999.9 sec
  • Ramp-Down: 1.0 - 999.9 sec (0=OFF)
  • Accuracy: ± (0.1% of reading + 0.05 sec)

Delay Timer 

  • Range: 0, 0.5 - 999.9 sec (0 = Constant)
  • Accuracy: ± (0.1% of reading + 0.05 sec)

Ground Fault Interrupt

  • GFI Trip Current: 450 μA max
  • HV Shut Down Speed: < 1 ms

General Specifications

Remote Control & Signal Output

The following input and output signals are provided through two 9 pin D type connectors:

1. Remote control: Test, Reset, and Remote Interlock

2. Remote recall of memory program #1, #2, and #3

3. Outputs: Pass, Fail, Test-in-process, and Reset

  • Program Memory: 10 Memories, 3 steps per memory
  • Security: Key Lock capability to avoid unauthorized access to all test parameters and memory locations
  • Mechanical: Bench or rack mount with tilt up feet
  • Dimensions: (W x H x D) 8.46 x 3.5 x 14.57 in. (215 x 89 x 370 mm)
  • Calibration: Traceable to National Institute of Standards & Technology (NIST)
  • Calibration controlled by software
  • Adjustments are made through front panel keypad in a restricted access calibration mode
  • Calibration information stored in non-volatile memory
More Information
ManufacturerAssociated Research
ConditionUsed
Looking for other manuals? Search all of our available manuals here.