Features:
- Large target Dimensions of 152mm (6") captures "peripheral vision" of infrared thermometers and minimizes
- Dimensions of source effect errors
- Compensates for errors caused by individual infrared thermometer emissivity settings
- Sample calibration procedures for Fluke and Raytek thermometers 3are built right in
- Accuracy as good as ±0.35°C
- Test Uncertainty Ratios (TUR) as good as 4:1
- Excellent stability and uniformity
- Reaches temperature setpoint quickly
- Intuitive, easy-to-read display that indicates when temperature is stable
Specifications:
- Display accuracy: ± 0.40 °C at –15 °C; ± 0.40 °C at 0 °C; ± 0.50 °C at 50 °C; ± 0.50 °C at 100 °C; ± 0.55 °C at 120 °C
- Stability: ± 0.10 °C at –15 °C; ± 0.05 °C at 0 °C; ± 0.10 °C at 120 °C
- Uniformity (5.0 in dia of center of target): ± 0.15 °C at –15 °C; ± 0.10 °C at 0 °C; ± 0.25 °C at 120 °C
- Uniformity (2.0 in dia of center of target): ± 0.10 °C at -15 °C; ± 0.10 °C at 0 °C; ± 0.20 °C at 120 °C
- Heating time: 15 min: –15 °C to 120 °C; 14 min: 23 °C to 120 °C
- Cooling time: 15 min: 120 °C to 23 °C; 20 min: 23 °C to –15 °C
- Dimensions (HxWxD): 356 mm x 241 mm x 216 mm (14 in x 9.5 in x 8.5 in)
- Safety: EN 61010-1:2001, CAN/CSA C22.2 No.61010.1-04
- Target diameter: 152.4 mm (6 in)
- Computer interface: RS-232
The Fluke 4180-156 is a temperature calibrator. It provides calibration accuracy and traceability for infrared thermometers and thermal imagers. Emissivity is radiometrically calibrated for consistent, accurate and traceable calibrations.
Manufacturer | Fluke |
---|---|