4181-156 Fluke Temperature Calibrator
New
$9,851.00
Features:
- Large target Dimensions of 152mm (6") captures "peripheral vision" of infrared thermometers and minimizes
- Dimensions of source effect errors
- Compensates for errors caused by individual infrared thermometer emissivity settings
- Sample calibration procedures for Fluke and Raytek thermometers 3are built right in
- Accuracy as good as ±0.35°C
- Test Uncertainty Ratios (TUR) as good as 4:1
- Excellent stability and uniformity
- Reaches temperature setpoint quickly
- Intuitive, easy-to-read display that indicates when temperature is stable
Specifications:
- Temperature range (@ 23°C ambient, 0.95 emissivity) - 35°C to 500°C
- Display accuracy1 - ±0.35°C at 35°C, ±0.50°C at 100°C, ±0.70°C at 200°C, ±1.20°C at 350°C, ±1.60°C at 500°C
- Stability - ±0.05°C at 35°C, ±0.20°C at 200°C, ±0.40°C at 500°C
- Uniformity (5.0 in dia of center of target)2 - ±0.10°C at 35°C, ±0.50°C at 250°C, ±1.00°C at 500°C
- Uniformity (2.0 in dia of center of target)2 - ±0.10°C at 35°C, ±0.25°C at 250°C, ±0.50°C at 500°C
- Heating time - 20 min: 35°C to 500°C
- Cooling time - 100 min: 500°C to 35°C, 40 min: 500°C to 100°C
- Stabilization time - 10 minutes
- Nominal emissivity3 - 0.95
- Thermometer emissivity compensation - 0.9 to 1.0
- Target diameter - 152.4 mm (6 in)
- Computer interface - RS-232
- Power - 115 V AC (±10%), 10 A, 50/60 Hz, 1000 W, 230 V AC (±10%), 5 A, 50/60 Hz, 1000 W
- Fuse(s) - 115 V AC 10 A, 250 V, fast blow, 230 V AC 5 A, 250 V, F
- Size (H x W x D) - 356 x 241 x 216 mm (14 x 9.5 x 8.5 in)
The Fluke 4181-156 is a temperature calibrator. It provides calibration accuracy and traceability for infrared thermometers and thermal imagers. Emissivity is radiometrically calibrated for consistent, accurate and traceable calibrations.
Manufacturer | Fluke |
---|---|
Condition | New |