19055-C Chroma HiPot

New

Additional Features:

  • Hi-Pot: AC 5kV/100mA and DC 6kV/25mA
  • Insulation: 5kVmax and 1MΩ~50GΩ
  • 500VA output rating
  • Floating output complies with EN50191
  • Corona Discharge Detection (CDD, 19055-C)
  • Flashover Detection
  • Discharge Level Analysis (DLA)
  • Open Short Check (OSC)
  • High Frequency Contact Check (HFCC)
  • Ground Fault Interrupt
  • Standard RS-232 interface
  • Option GPIB & HANDLER interface
  • Key lock when fail
  • Programmable voltage & test limit
  • CE Mark

The 19055-C has not only the AC/DC/IR tests but also a new measurement technology - Corona Discharge Detection (CDD) that can detect the following via the Discharge Level Analysis (DLA):

  • Corona discharge Start Voltage (CSV) 
  • Flashover Start Voltage (FSV) 
  • BreakDown Voltage (BDV)

The 19055-C HiPot Analyzer with 500VA output rating can be used to test and analyze the withstand voltage of high power and leakage current for the products like motor stators and rotors with high parasitic capacitance. Corona detection can be used for turn-to-turn or turn-to-ground test to avoid winding insulation failure from corona discharge.

When using a power transformer under the normal voltage, a primary side corona discharge could cause the adjacent components to be damaged if occurred. Thus, the function of Corona Discharge Detection (CDD) of 19055-C can be used to detect if there is any corona discharge occurred to improve the product quality.

More Information
ManufacturerChroma
ConditionNew
No manuals are currently available. Search all of our available manuals here.

Accessories